Via Pellizza da Volpedo 101A,
 20092 Cinisello B. (MI)
Automatic Isokinetic Sampler

ST5

The reliable automatic sampler every isokinetic Source Tester needs

Take the worry out of isokinetic sampling and eliminate the risk of human error in manual data

entries and calculations. Let’s make your work easier and more accurate!

When paired with the Heated Probe HP5, Cold Probe CP5 and CP2, the ST5 stands out as the most advanced technical solution available on the market.

Meets and exceeds EN standards EN 16911-1, EN 15259, EN 13284, EN 1948-1, OTM 45

Compatible with US EPA Methods M2, M5 ed M17

Works with US EPA probes.

Operating range 5 ÷ 60 l/min
Sampled gas condition Dehumidified, max temperature 45°C
Gas inlet With particulate and liquid protection filter
Connections Quick couplings
Operating conditions -20 ÷ 45°C, 95% RH
Storage conditions -10 ÷ 50°C, 95% RH
Display 3.5” Graphic LCD (QVGA)
Data port USB 2.0, Bluetooth
Internal memory 16 GB
Power supply 230 Vac ±10% 50/60Hz
Materials Structure composed of steel and aluminum
Keyboard Polycarbonate, tactile effect keys
Dimensions ST5 390 × 264 × 506 mm
Dimensions ST5 EVO CU 390 × 264 × 386 mm
Dimensions ST5 EVO PU 227 × 329 × 191 mm
Weight ST5 pump 4.5 m³/h 22 kg
Weight ST5 pump 8.5 m³/h 24 kg
Weight ST5 EVO CU 13 kg
Weight ST5 EVO PU pump 4.5 m³/h 10 kg
Weight ST5 EVO PU pump 8.5 m³/h 11.5 kg
Measurement and sensor characteristics
Sampled volume
DGM Class G2.5 EN 1359 compliant
Flow range 0.025 m³/h ÷ 4.0 m³/h
Accuracy 2% reading
Encoder resolution 0.1 liters
Sampling flowrate
Control type Orifice mass flow meter, compliant with UNI EN ISO 5167-2
Range 5 ÷ 60 l/min
Resolution 0.01 l/min
Accuracy ± 1%
Pitot differential pressure
Range -100 ÷ 1000 Pa (-10 ÷ 100 mmH₂O)
Linearity and hysteresis (combined) 0.25 % FS
Accuracy Better than 1% of measurement (± 2 Pa)
Resolution 0.01 Pa (0.005 mmH₂O)
Maximum burst pressure Max 3000 Pa (300 mmH₂O)
Absolute pressure (static-barometric)
Range 10 ÷ 105 kPa (1050 mBar)
Linearity and hysteresis (combined) 0.25 % FS
Resolution 0.01 kPa (0.1 mBar)
Accuracy Better than 1% of measurement (± 0.25 kPa)
Thermocouple input
Range -20 ÷ 1200 °C
Resolution 0.1 °C
Accuracy 1% reading ± 0.4 °C
DGM temperature
Range -20 ÷ 100 °C
Resolution 0.1 °C
Accuracy 1% reading ± 0.2 °C